Full-Spectrum Direct-Reading Spectrometer SparkCMOS 8000

2026-04-24

The working curve in the standard configuration of this instrument is plotted using standard samples by Steel Research NACK Jiangsu Testing Technology Research Institute Co., Ltd. For special alloys and elements, the user should provide standard samples or control samples with defined values, which NACK will plot for free.  

3. SparkCMOS 8000 Spectrometer Technical Specifications  

3.1 Instrument Appearance  

The instrument features a brand-new design, more ergonomically friendly, making operation quick and convenient.  

3.2 Optical System  

Full spectrum analysis  

Paschen–Runge mount method, grating focal length 500 mm  

High-emission holographic grating, 2700 lines/mm  

Wavelength range: 130 ~ 800 nm  

Resolution: better than 0.01 nm  

CMOS detector, no limit on maximum detection channels  

3.3 Brand-New Monitoring System  

The system monitoring is integrated at the bottom of the software interface, clear and easy to understand. It displays status, information, working curves, etc., fully showing instrument status and making maintenance and debugging easier.  

3.4 Control and Data Acquisition Electronic Unit  

16-bit analog-to-digital conversion microprocessor  

3.5 Spark Excitation Source  

Discharge parameters are password protected  

Fully digital plasma spark source, high-energy pre-spark technology, up to 1000 Hz  

Parameters such as source frequency and energy are continuously adjustable  

3.6 Integrated Acquisition  

Data acquisition via network port, stable data transmission, low computer requirements  

Multi-threaded data acquisition improves software stability and data reliability  

Up to 12 data acquisition channels, covering a wide range  

Uses segmented integration method for acquisition, with 12 selectable exposure levels, increasing flexibility  

Acquisition board uses lead-free solder, reducing pollution and hazards

3.7 Spark Platform

The spark stage measures 125mm × 95mm, with samples weighing up to 50Kg, and is radiation-free.

The newly designed coaxial spark stage uses an optimized internal air path, greatly reducing argon consumption, minimizing residual metal dust inside the spark stage, and improving the stability of the instrument when analyzing data

Integrated lens isolation valve: the lens and isolation valve are integrated, preventing strength loss caused by optical chamber contamination caused by daily maintenance, and the lens is easy to replace

The discharge chamber is uniquely designed to ensure discharge under optimal conditions

The excitation electrode is the tungsten electrode

3.8 Argon-filled optical system

The argon filling system is controlled by valve islands and software, enabling rapid argon filling and low air consumption

A newly designed cast optical chamber with an extremely low coefficient of thermal expansion greatly enhances the instrument's stability

The brand-new constant flow argon filling system is stable and reliable, ensuring long-term stability of the optical system and reducing spectral line drift

3.9 Analysis Software

Including calculations for different substrates and curves

Multilingual analysis software for easy user operation

The software allows the following operations:

Ø Establish and modify the analysis program

Ø Selectable output methods (intensity, intensity ratio, uncalibrated concentration, corrected concentration, etc.)

Ø Concentration units for analysis results can be selected

Ø Universal standardization is possible

Ø Standardization of types

Ø Standardization of multiple sample types

Ø Corrects interference between elements

Ø Can print all results from over 100 excitations, along with their averages, standard deviations, and relative standard deviations

Ø Excitation can be repeated once or multiple times at the same point, and the result can be removed once or multiple times

Ø Standard Material Library

Ø Statistical analysis and calculations can be performed

Ø Output results in different languages

Ø The analysis results can be output in formats acceptable to Office software

Ø Hard drive data storage facilitates future research

Ø Can connect to external computers or central systems

3.10 Power Requirements

220V single-phase 10A 2.5KVA

Ensure the power supply is a sine wave; Voltage fluctuation less than ±10%; Frequency fluctuation less than ±2%

3.11 Dimensions

Width× depth× height (54cm×86cm×48cm)

3.12 Weight

Net weight is about 100 kg

Gross weight is about 120 kg

3.13 Work environment

Argon purity: not less than 99.999%

Laboratory temperature range: 20 ~ 25°C

Laboratory humidity: relative humidity less than 70%

Ground wire: dedicated ground wire, grounding resistance less than 4 ohms

Shock resistance: Shock resistance is 10 accelerations of gravity, and the laboratory floor shows no noticeable vibration

Electromagnetic field interference: RF signal attenuation greater than 103, and no medium-frequency furnaces, transformers, or other high-frequency generation devices near the laboratory

Dust requirements: less than 500 particles per kiloliter